Impedance‐Based Contact Visualization for PulseSelect Pulmonary Vein Isolation: A Single‐Center Feasibility Study
Year: 2026
Type: article
Abstract: ABSTRACT Background PulseSelect pulmonary vein isolation (PVI) lacks a dedicated contact indicator. Methods We retrospectively compared 63 patients undergoing PVI before and after EnSite Contact Index (CI) implementation. Because electrical isolation does not confirm lesion transmurality, post‐procedural unipolar electrograms (u‐EGMs) were assessed as acute surrogate markers. Results CI guidance was associated with more applications, shorter fluoroscopy, and in an exploratory analysis of analyzable patients, fewer residual high‐voltage u‐EGMs (1/20 vs. 8/24; p = 0.027). No complications occurred. Conclusions CI‐guided PVI was feasible; these procedural and acute electrophysiological associations require prospective validation.
Source: Journal of Arrhythmia
DOI: 10.1002/joa3.70507
Institution: Kyoto Katsura Hospital
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $1,540
OpenAlex ID: W7225614398
