Abstract: Physics experiments at SLAC require high accuracy positioning, e. g. 100 {micro}m over a distance of 150 m or 25 {micro}m in a 10 x 10 x 3 meter volume. Laser Tracker measurement systems have become one of the most important tools for achieving these accuracies when mapping components. In order to improve and get a better understanding of laser tracker measurement tolerances we extended our laboratory with a rotary calibration table (Kugler GmbH) providing an accuracy of better than 0.2 arcsec. This paper gives an overview of the calibration table and its evaluation. Results of tests on two of our Laser Trackers utilizing the new rotary table as well as the SLAC interferometer bench are presented.
Authors: Georg Gaßner, R. Rüland
Language: English
Subfield: Mechanical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: green
Funder: U.S. Department of Energy
Awards: G2371548206, G303061748, G5570207115