Charging: a Limitation to Perform X-ray Microanalysis in the Variable Pressure Scanning Electron Microscope

Year: 2005
Type: article
Abstract: The variable pressure scanning electron microscope (VP-SEM) allows the imaging of insulators without coating.The electrons, resulting from the interaction between the incident beam and the specimen, interact with gas molecules.Some positive ions are created and they move towards the surface of the specimen in order to neutralise it.Charging (positive or negative) occurs when a material is not able to conduct effectively the beam energy imparted to it.Charging is a phenomenon which is not perfectly known, due to the difficulty to evaluate it.Nevertheless, several studies have been published and some techniques have been proposed to evaluate this phenomenon.A measure of the specimen charging can be performed by measuring the Duane-Hunt limit (DHL) [1].